- ページ 2

ラボラトリー機器 Distek 2500 SelectのPDF サービスノートをオンラインで閲覧またはダウンロードできます。Distek 2500 Select 6 ページ。
Distek 2500 Select にも: クイック・スタート・マニュアル (10 ページ)

Distek 2500 Select サービスノート
Before proceeding, discharge any static electricity from your body by touching a bare screw from the
instrument. Turn off the instrument and disconnect the AC power cord.
Solution / Action
Using a Phillips Head Screw Driver, remove the nine screws from underneath the drive plate and
the one screw located on the top rear of the cover.
Step 1
Remove the drive plate cover by holding both sides and raising the cover up.
Step 2
Remove the spindle drive belt by pulling it off the idler wheel as shown.
Step 3
Distek, Inc.
121 North Center Drive
North Brunswick, NJ 08902 USA
P: 732-422-7585 F: 732-422-7310
This document contains confidential and proprietary information which is not to be disclosed to others for any purpose nor used for manufacturing
purposed with written permission from Distek, Inc.
Service Note:
TOLL FREE: 888-234-7835
Service Note
Step 1
Step 2
Step 3
2500x-1090
Page 2 of 6