Tip Cleaning & Sterilization:
The included tips have not been sterilized and must be sterilized prior to use.
The tip of the Heron IOS is inserted into the patient's mouth during operation.
To prevent cross-contamination between patients the tip should be cleaned and
sterilized according to the following procedure.
Step 1: Clean the tip with soap water ensuring that the mirror is clean and free of
smudges, stains, or any milky haze. Do not use an abrasive surface to clean
the mirror.
Be careful when cleaning
the mirror, it is delicate
and prone to scratching.
Step 2: After drying the tip exterior, carefully dry the interior and mirror with
delicate task wipes (such as KimWipes®) ensuring you do not scratch the
surface. The mirror should be free of any noticeable blemishes prior to
continuing.