socket at the rear of the unit and secure probe into the holder. Turn the
instrument on.
3. The temperature display will show the actual temperature of the sample as
measured by the P1000 probe.
Maintenance and Cleaning
• Proper maintenance and cleaning can keep instruments working properly and
will lengthen its lifetime.
• ABS casing will give some chemical resistance to weakly acidic and alkaline
liquids.
• Do not spray your chosen cleaner directly on to the unit when cleaning
• Ensure the unit is unplugged from the power supply and cooled down
completely before commencing with cleaning.
• Wear appropriate PPE equipment during the cleaning process.
• Only use recommended cleaners below:
Dyes Isopropyl alcohol
Construction
materials
Cosmetics
Foodstuffs Water containing surfactant
Fuels Water containing surfactant
• If the instrument will not be in operation for an extended period of time, please
switch off and store in a dry, clean environment.
Faults and Troubleshooting
Instrument is not powering on
• Check whether the power line is plugged in
• Check whether the fuse is broken or loose
Stir speed is not reaching set point
• Excessive medium viscosity may cause abnormal speed reduction of the
motor. This instrument has only been verified for stirring water and mediums
with a similar viscosity). Try the unit with a small quantity of water to check the
Water containing surfactant/
isopropyl alcohol
Water containing
surfactant / isopropyl alcohol