30471PA, CAP. SENSOR TESTER KIT INSTRUCTIONS
6. Turn on the capacitance sensor tester.
7. Verify the capacitance sensor tester shows closed.
8. Begin unscrewing the knob.
9. Verify the capacitance sensor tester switches from
Replacement Parts
9V Battery: Dixon 10261
Storage Case: Dixon 10491
REV A 2018/06/11
If the capacitance sensor tester shows any of the
following conditions, the sensor is considered bad:
a. Closed
b. No sensor
c. Shorted
Removed from service and contact Dixon technical
support.
closed to open before the knob is completely
unscrewed. Be sure to remove your hand from the
knob as this will interfere with the gauges
accuracy. If the knob is completely unscrewed and
the sensor output tester shows any of the
following conditions, the sensor is considered bad:
a. Closed
b. No sensor
c. Shorted
Remove from service and contact Dixon technical
support.