Symbol Definitions
Power on/off button and alarm reset
S Mode Operation
Compliance monitor reset
1. Set up Pump
Press bed hanger release on the back of the pump and gently pull out
bed hanger hook to the desired width.
Connect tubing to the pump. Plug pump into electrical outlet.
2. Apply Cuff
The VenaFlow Elite system will automatically detect the type of cuff and the number of cuffs attached and apply appropriate
pressures. Cuffs must be attached prior to powering on the service.
Be sure to secure cuff straps snug but not tight. When using the calf cuff and thigh cuff, rotation of the aircell on the leg does not
affect performance of the system.**
Calf or Thigh Cuff:
For varied patient sizes, trim for proper, snug fit.
Apply cuff snugly with the tube pointed
toward the foot. Aircell can be orientated to
patient and nursing preference (posterior,
lateral, anterior, or medial).**
1
7
2
6
3
5
4
Tube alarm
Call for service
Inflation cycle indication
(Bars indicate cuffs are inflating)
C
Visual only compliance alarm enabled
Inflation Cycle
5
indicator
Patient Compliance
6
counter
Patient Compliance
7
counter reset push button
ON/OFF/RESET
8
push button
See VenaFlow Elite System
operator's manual
Battery
Lock mode
Visual and audio compliance
alarm enabled
Foot Cuff:
Apply foot cuff snugly with the aircell
centered on bottom of foot and with the
tube pointed to the left. Foot cuffs are only
for use in standard rapid mode.
S Mode Operation
button
Battery Indicator
*Note: Not all VenaFlow Elite
Systems contain a battery
Single/Dual Leg
operation indicator
S Mode Operation
indicator
Dual calf/thigh
Single calf/thigh
Dual foot
Single foot
Bed hanger release
Graphical display
Pump indicator lights
(green = on, flashing
red = alarm)
Telescoping bed
hanger hook