- ページ 3

ストレージ CUSTOM ULTRASONICS 20000のPDF 取扱説明書をオンラインで閲覧またはダウンロードできます。CUSTOM ULTRASONICS 20000 11 ページ。 Endoscope storage cabinet
CUSTOM ULTRASONICS 20000 にも: 取扱説明書 (7 ページ)

 
 
CUSTOM ULTRASONICS, INC. 

WARNING / CAUTIONs

A WARNING is a statement that alerts the user to the possibility of injury, death, or other serious adverse
reactions associated with the use or misuse of the Storage Cabinet.
A CAUTION is a statement that alerts the user to the possibility of a problem with the Storage Cabinet
associated with its use or misuse. Such problems include product malfunction, product failure, damage to
the product or damage to other property.
WARNING: For installation of the Storage Cabinet, make certain to have a minimum of two (2) persons to
support and mount the device, otherwise, bodily injury and physical damage to the Storage
Cabinet and surrounding area may occur.
WARNING: The Storage Cabinet will tip over if not properly secured.
WARNING: Ensure the Storage Cabinet's electrical plug is inserted into a properly grounded and
dedicated, 120 VAC, 15-20A, 60 Hz, 1 receptacle. Receptacle must be accessible to the
operator for quick disconnect.
CAUTION:
Follow all Automated Endoscope Washer/Disinfector and endoscope manufactures
instruction for thorough drying of the endoscopes prior to placing into the Storage Cabinet.
CAUTION:
Follow endoscope manufactures instruction for handling and storage of an endoscope.
CAUTION:
The Storage Cabinet is designed to store up to eight (8) endoscopes maximum. Do not
attempt to store more than the Storage Cabinet's intended capacity.
CAUTION:
One adapter must remain in the adapter manifold at all times to prevent damage to the air
pump.
CAUTION:
Make certain the endoscopes are not hanging out of the Storage Cabinet. This could result
in damage to the endoscope and/or Storage Cabinet when closing the door.

Labels and Label Indications

Label 
MAN-009
Graphical Label: Indicates the Air Purge Set Time (0 to 30 minutes).
CAUTION: The label indicates to disconnect from power source prior
to removing or opening door.
French version of the label is above.
WARNING Symbol: Attention: Potential shock hazard.
CAUTION Label: Ground reliability can only be achieved when the
device is connected to an equivalent receptacle mark hospital grade.
Revision 08232021
Endoscope Storage Cabinet
Model 20000
Operator's Manual
Indication
Page 3 of 11