- ページ 12
ラボラトリー機器 DLab MS7-H550-SのPDF ユーザーマニュアルをオンラインで閲覧またはダウンロードできます。DLab MS7-H550-S 19 ページ。 7’square (hotplate) magnetic stirrer
operate follow the instructions below:
•
Switch OFF the instrument.
•
Ensure the external temperature sensor is inserted in
the media heated.
•
Switch ON the instrument and run heating function.
If the heating function does not work, please contact
manufacturer/supplier for technical support.
7. Residual heat warning(HOT)
In order to prevent the risk of burns from the hotplate,
digital hotplate model has a residual heat warning function.
When the heating function is switched off and the heating
plate temperature is still above 50°C, "Hot" will flash to
warn that there is a hazard of burns from the hotplate. When
the hotplate temperature drops to below 50° C, the unit will
automatically switch off. If users want to turn off the LCD
or LED screen immediately, just pull out the plug directly.
When the power OFF, the residual heat warning function
cannot be run.
8. Remote Control (LCD Digital
model)
The unit can be controlled from an external PC (using the
dedicated software) via the RS232C serial interface fitted to
the unit. Data communication from laboratory instrument to
computer is only possible on demand of the computer.
•
The functions of the interface lines between laboratory
instrument and automation system are selected from
the specified signals of the EIA-standard RS232C,
corresponding with DIN66020 Part 1. The allotment of
the bushing can be taken from Figure 8.
•
Transmission
method:
10
7" Square Hotplate Magnetic Stirrer
Figure 8
Asynchronous
signal