Symptom
The sling does not fit the user.
All the straps do not lift at the same time.
User is sliding through/out of the sling.
Troubleshooting
Possible Cause/Action
Wrong sling size. Try other size.
The sling is too small when the leg supports do
not reach halfway around the inner thigh and
the suspension is too close to the user's face. Try
other size.
The sling is too small when top of sling doesn't
support the users head and the suspension is too
close to the user's face. Try other size.
The sling is not correctly positioned.
Pull the sling further down at the back when
fitting it.
If the sling straps lift unevenly, it means that the
sling can slide up at the back which can lead to
the user sliding down in the opening of the sling.
Make sure that the leg supports are positioned
correctly on the thighs.
The sling is not connected correctly according to
individual prescription.
The user is wearing to slippery clothes. Change or
remove some of the clothing