ラボラトリー機器 Etaluma LS460のPDF スタートアップマニュアルをオンラインで閲覧またはダウンロードできます。Etaluma LS460 14 ページ。 Lumascopes with lumaview (for use with ls560, ls460, ls600, ls620, ls720)

2. Connect the free end of the Phase Accessory communication cable to
the round port on the LS560 back panel (left side) labeled External
Lamp Power Out. This allows the Phase Accessory to be controlled by
Lumaview.
F. Brightfield
1. For optimal brightfield, use of the Phase Accessory (Section I.E. above) with the included Slider in an
open position (no phase ring) is recommended. The Phase Accessory provides collimated light and is
directly connected to the LS560, allowing it to be controlled by Lumaview. It should also be used when
imaging in a dark environment such as an incubator. To mount the Phase Accessory above the LS560
deck, see Section I.E. above. Brightfield images will be in gray scale due to the monochromatic CMOS
camera.
2. If the Phase Contrast Accessory is not available, ambient light in a typically lit area can sometimes be
used. The light should be as uniform as possible to prevent shadows or other dark or light areas.
Overhead fluorescent lighting can cause uneven brightness including a striped pattern; if observed,
move the LS560 or partially shade the light to reduce unevenness across the sample.
G. LED Power (not applicable to the LS460)
For almost all fluorescence imaging applications, power from the computer via the USB cable to the
LS560 will be sufficient. If high levels of illumination are required, e.g., with very dim samples, plug the
small round plug of the external power supply/cord into the round port labeled SUPPLEMENTAL
POWER on the back of the LS560 and the electrical plug into an AC outlet. You should see a moderate
increase in the Illumination level.
H. Installing Objectives
1. Manually remove the insert of the accessible deck. Unscrew the black cap (or previously installed
objective) on top of the optics block and screw in the new objective. Do not overtighten. Return the
removable insert to the deck.
2. When installing or changing an objective, carry out the replacement quickly to minimize the chance of
dust falling onto the mirrored dichroic filter. If any dust is observed on the filter, carefully blow it off
using a compressed air duster, making sure to hold the can UPRIGHT and bending the straw to aim the
air directly toward the filter. DO NOT tilt the can or liquid can be emitted, which will require further
cleaning. Other methods to remove dust may damage the dichroic filter and must NOT be used.
I.

About Lumaview

1. The LS560 is controlled by the Lumaview software program. The latest Lumaview version is
downloadable from Etaluma's website and must be installed prior to connecting your computer for
the first time to the LS560 (due to the need for two Drivers that are installed on the computer at the
same time).