SETTING UP THE UNIT:

FIG. 1 - ADU-17A Express Air
HANDPIECE
HIGH VOLUME VACUUM
(HVE)
COOLANT ON/OFF
SWITCH
AIR CONTROL
VALVE
1.
Unpack the ADU-17A Case.
2.
Verify that the Voltage Selector Insert is configured properly
by checking the voltage indicator on the Voltage Selector
Module (Figure 2a). The Voltage Module is located on the
lower left of the Motor Cover. The Voltage Selector Insert
must be installed in the "120" position for 120V 60Hz
operation or in the "240" position for 230V 50Hz operation.
ALWAYS DISCONNECT POWER TO THE SYSTEM BEFORE
OPENING VOLTAGE SELECTOR MODULE
To switch the voltage configuration, disconnect power cord
and insert a small flat screwdriver under the locking tab in the
front of Selector Module. Gently depress tab until the fuse
housing releases (Figure 2b). Carefully remove fuse housing
to reveal Voltage Selector Insert (Figure 2c). Pull insert out of
housing, turn it around and reinsert into housing with proper
voltage indicator facing out (Figure 2d).
4.
WASTE TANK TIE-
DOWN
CAVITRON
CONNECTOR
COOLANT
CONTROL VALVE
WARNING:
3-WAY AIR/WATER SYRINGE
CLIP-ON INSTRUMENT HOLDER
LOW VOLUME VACUUM (SALIVA
EJECTOR)
RESEVOIR CONNECTOR
WATER SUPPLY BOTTLE
COMPRESSOR AND FAN
HOUSING
ON/OFF SWITCH
VOLTAGE SELECTOR
MODULE
IMPORTANT: For 120V use, check that the Fuse in use
is a 6A medium time delay; for 230V use, check that
the Fuse is a 4A Slo-Blo type. User may have to change
fuses before use.
For 230V use, confirm that the power cord is correct for the
country of usage and carries the proper certifications.
Carefully reinstall fuses into Fuse Holder and slide Holder
back into Selector Module (Figure 1e). Apply pressure until
locking tab snaps back into position, producing a distinct
clicking sound.
3.
Mount the clip-on Holder to the extended handle of the Case,
as shown in Figure 3.
CASE HANDLE
WASTE TANK
FOOT PEDAL
FUSES LOCATED IN
VOLTAGE SELECTOR
MODULE