INSTRUCTIONS FOR USE - HANDPIECES
For trained professional use only
Indication for use:
Implant dentistry including; perforating bone, tapping and threading implants
AHP-64, AHP-85I, AHP-85P-I, AHP-85AS-I, AHP-85MB-X, AHP-85MB-CX,
AHP-85MBFO-CX
For endodontic dentistry; drilling into the root canal
AHP-62M, AHP-62MP, AHP-88, AHP-88MP,
For general dentistry; removing a carie from the dentine
AHP-63, AHP-64
Disinfection & Sterilization Procedures:
Disinfect the surface only. Never immerse the entire handpiece in disinfectant
solutions. Disinfecting the surface does not provide adequate infection control
between patient contact.
This dental handpiece, external tubing and clips, and the above referenced
handpieces, are autoclavable and must be steam autoclaved before initial use
and between each patient use per the following procedure:
1)
Scrub dirt and other debris from the handpiece and wipe clean with a cotton
swab or cloth.
2)
Lubricate with Aseptispray or other handpiece lubrication approved for this
application.
3)
Autoclave at 135°C (275°F) for 20 minutes.
Warranty:
One (1) year against defective parts or workmanship, from date of original invoice.
Accessories:
Handpiece tubing
(may not be included with all handpiece models)
1. FOR EXTERNAL IRRIGATION: Attach small diameter
tubing to contra angle head on tube provided under
the chin of the handpiece head.
2. FOR INTERNAL IRRIGATION: Attach 2 clips to large
diameter tubing and attach to handpiece as shown.
Insert cannula into hole provided in top of head and
push away. Small clip will fit to handpiece behind latch.
3. Chrome 1/16" connector; for connecting either internal
irrigation or external irrigation to pump. Both irrigators
cannot be used at same time with this connector.
4. Y-connector; for connecting external and internal
irrigation for simultaneous use. The irrigation will take
the line of least resistance, so the bulk of the fluid will
flow to the external tubes.
8333 - 216th Street SE • Woodinville, WA 98072
Phone: (425) 487-3157 • Toll Free: (800) 426-5913 • FAX: (360) 668-8722
www.aseptico.com • [email protected]
For trained professional use only
Indication for use:
Implant dentistry including; perforating bone, tapping and threading implants
AHP-64, AHP-85I, AHP-85P-I, AHP-85AS-I, AHP-85MB-X, AHP-85MB-CX,
AHP-85MBFO-CX
For endodontic dentistry; drilling into the root canal
AHP-62M, AHP-62MP, AHP-88, AHP-88MP,
For general dentistry; removing a carie from the dentine
AHP-63, AHP-64
Disinfection & Sterilization Procedures:
Disinfect the surface only. Never immerse the entire handpiece in disinfectant
solutions. Disinfecting the surface does not provide adequate infection control
between patient contact.
This dental handpiece, external tubing and clips, and the above referenced
handpieces, are autoclavable and must be steam autoclaved before initial use
and between each patient use per the following procedure:
1)
2)
3)
Warranty:
One (1) year against defective parts or workmanship, from date of original invoice.
Accessories:
Handpiece tubing
1. FOR EXTERNAL IRRIGATION: Attach small diameter
2. FOR INTERNAL IRRIGATION: Attach 2 clips to large
730298
TUBING
3. Chrome 1/16" connector; for connecting either internal
4. Y-connector; for connecting external and internal
INSTRUCTIONS FOR USE - HANDPIECES
Scrub dirt and other debris from the handpiece and wipe clean with a cotton
swab or cloth.
Lubricate with Aseptispray or other handpiece lubrication approved for this
application.
Autoclave at 135°C (275°F) for 20 minutes.
(may not be included with all handpiece models)
tubing to contra angle head on tube provided under
the chin of the handpiece head.
diameter tubing and attach to handpiece as shown.
Insert cannula into hole provided in top of head and
push away. Small clip will fit to handpiece behind latch.
irrigation or external irrigation to pump. Both irrigators
cannot be used at same time with this connector.
irrigation for simultaneous use. The irrigation will take
the line of least resistance, so the bulk of the fluid will
flow to the external tubes.
8333 - 216th Street SE • Woodinville, WA 98072
Phone: (425) 487-3157 • Toll Free: (800) 426-5913 • FAX: (360) 668-8722
www.aseptico.com • [email protected]
730298
TUBING