VIII. TECHNICAL SPECIFICATIONS

8.0
OUTPUT CHARACTERISTICS
Mode/
Output
Character
Waveform
Cut
4.0 MHz
sinusoid
Cut/Coag
4.0 MHz with
fully rectified
envelope
Hemo
4.0 MHz with
partially rectified
envelope
Bipolar
4.0 MHz with
Hemo
partially rectified
envelope
RF Power-out is within 20% of nominal above (when the dial setting is greater than 1).
RF Power-out is stable within 10% within 10 seconds after activation.
RF output frequency is within 10% of nominal.
NOTE: The open-circuit output peak-to-peak voltage for all modes is approximately
1200 Volts.
8.1
PREVENTIVE INSPECTION AND MAINTENANCE
Regular preventive inspection should be carried out to prevent reduced safety of the unit due to
aging, wear and tear, etc. The cable and electrode must be checked before each use.
The Generator must be inspected for technical safety at least once a year. The required technical
safety inspection is detailed in the Maintenance Manual.
The manufacturer assumes no responsibility for improper changes or repairs carried out on the
unit or its accessories by unauthorized persons, and the warranty of the unit expires immediately.
Nominal
Power Output
Output Power
vs. Setting
50W
See Figure 7.0
@ 500 Ω
50W
See Figure 7.0
@ 500 Ω
35W
See Figure 7.0
@ 500 Ω
35W
See Figure 7.0
@ 200 Ω
17
Power Output
Activation
vs. Load Resistance
See Figure 7.1
via footswitch or
fingerswitch
See Figure 7.2
via footswitch or
fingerswitch
See Figure 7.3
via footswitch or
fingerswitch
See Figure 7.4
via footswitch