- ページ 16

ラボラトリー機器 Clontech PT3139-1のPDF ユーザーマニュアルをオンラインで閲覧またはダウンロードできます。Clontech PT3139-1 20 ページ。 Advantage-hf pcr kit

CLONTECH Laboratories, Inc.
V. Troubleshooting Guide continued
Touchdown PCR
needed
Contamination
C. Products are smeared
Too many cycles
Denaturation temp. Try increasing the denaturation temperature in incre-
too low
Extension time
too long
Poor template
quality
Touchdown PCR
needed
Too much enzyme
2+
[Mg
] is too high
Too much template Try a lower concentration of DNA template in the
Contamination
page
Protocol # PT3139-1
16
Version # PR76834
"Touchdown" PCR significantly improves the speci-
ficity of many PCR reactions in various applications
(Don et al. , 1991; Roux, 1995). Touchdown PCR
involves using an annealing/extension temperature
that is several degrees higher than the T
primers during the initial PCR cycles. The annealing/
extension temperature is then reduced to the primer
T
for the remaining PCR cycles. The change can be
m
performed either in a single step or in increments
over several cycles.
See Section D below.
Reduce the cycle number by 3–5 to see if non-
specific bands go away.
ments of 1°C.
Decrease the extension time in 1–2-min increments.
Check template integrity by electrophoresis on a de-
denaturing agarose gel. Repurify your template if
necessary.
See "Touchdown PCR needed" under previous
section.
Advantage-HF Polymerase Mix is 50X for most appli-
cations; however, a 1X final concentration of the
enzyme mix may be too high for some applications.
If smearing is observed, first try optimizing the cycle
parameters as described above, then try reducing
the enzyme concentration to 0.5–0.2X.
KlenTaq-1 DNA polymerase has a broader Mg
optimum than native Taq DNA polymerase (i.e., it
performs well over a wider range of [Mg
loss of efficiency.) Therefore, as long as you have
used the buffers supplied in the kit, it is unlikely that
smearing is due to problems with the Mg
concentration. Altering the concentration of Mg
result in lower fidelity.
PCR reaction.
See Section D below.
Technical Support TEL:415-424-8222 or 800-662-CLON
FAX: 415-424-1064 or 800-424-1350
of the
m
2+
2+
] with no
2+
++
can