2.1
System Set-up
1. Turn the TTSM system on.
a. Ensure that the TTSM System is correctly connected to the computer (see
Section A.2).
b. Turn on the light source and ensure that it is set to 100% power.
2. Run Surface Measurement by clicking the Surface Measurement button in the
Studio Manager bar on the left side of the screen. A start-up splash screen appears
while the software initializes. Refer to the Surface Measurement User Manual for
operating procedures.
3. To access the sample tray, push firmly inward on the front of the sample drawer.
This will release the catch that holds the drawer closed, allowing the sample tray
to be pulled towards the user. Place all samples on the tray so that they are
oriented in line with the tray and parallel with each other.
4. Push the sample drawer closed until the catch engages. The system is now ready
to begin operation.
2.2
System Shutdown
The TTSM has a rocker switch on the back, near the main power cord, that serves
as the master on/off switch.
2.3
Replacing Gratings and Lenses
Both gratings and lenses can be swapped out to change the field of view, data
density, and Z-resolution requirements (see Table 1.1 for details). In order to swap
gratings, open the system doors and carefully lift out the old grating before putting a new
one in its place. To replace a lens, open the door and unscrew the old lens from the
camera, then screw on a new one.
2.4
Cleaning a Grating
Maintaining a clean grating is critical to optimizing measurement resolution and
reducing measurement noise when using shadow moiré. Gratings should be cleaned on
an as needed basis, when dust collects on the top surface or fingerprints are present.
Gratings should be cleaned with a lint free cloth or wipe and typically with a mild glass
cleaner or 30-70% IPA solution. The etched lines in the grating surface are not easily
removed. Users are free to press firmly on the grating and wipe in multiple directions to
clean the grating surface effectively. If the grating is removed for cleaning, be careful not
to drop it, as they are relatively fragile.
2 System Set-up and Use
Note: Refer to the cleaning product precautions for any
personal protection equipment needed and for disposal
of soiled cloths.
40-5000-8100 Rev. A
Page 9 of 14