- ページ 5

計測機器 Ashcroft 2074のPDF 取扱説明書をオンラインで閲覧またはダウンロードできます。Ashcroft 2074 13 ページ。 Digital industrial gauge

Ashcroft 2074 取扱説明書
KEYPAD FUNCTIONS
This key allows for selecting gauge features in the
menu finalizing selection. Use of the enter key is
ENTER
described throughout operating instructions.
MENU OPTIONS
UNITS: 12 units of measurement are available: psi, mmHG,
inH
O with three temperature options: 20°C, 60°F, 4°C*,
2
mBar, inHg, ftH
O, mPa, kPa, kg/cm
2
Step 1: Press the MENU key until the word UNITS appears.
Step 2: Press Enter.
Step 3: Press the
(up arrow key) or
key) to select the required unit of measure.
Step 4: Press ENTER to finalize the UNIT selection.
Note: For inH
0 range with three temperature options,
2
press the
(up arrow key) and
select the desired temperature, then press ENTER to final-
ize the UNIT selection.
CONFIG: This option allows access to additional Menu
options. Options include:
• ENTPW or enter password (this appears as a sub-menu
in the CONFIG mode if a user password has been set).
• RECAL (allows for zero, span and mid-scale calibra-
tion of the gauge).
• 0bUTN or zero key (allows for adjustment of % of
range that can be zeroed),
• dISAb, allows for disabling MENU options.
• SETPW: This feature allows for a user defined numer-
ic password. If a user password is not set, all features
in the CONFIG mode will be accessible without a
password. If a user password is set, all items in the
CONFIG menu options are accessible with or without
a user password. If a user password is programmed,
it will be required to access the CONFIG menu options.
*Indicates default.
– 8 –
2
and bar.
(down arrow
(down arrow key) to
KEYPAD FUNCTIONS
MENU FUNCTIONS
How to Use Your Menu Functions
To set a user password (SETPW):
Step 1: Press the Menu key on the keypad
Step 2: Press the
(up arrow key) or
key) until the word CONFIG appears.
Step 3: Press Enter. The word SETPW appears on the
gauge display
Step 4: Press Enter. A five digit numeric password is now
required.
Step 5: Press the
(up arrow key) or
key) on the keypad to select the first digit of the password.
Step 6: Press ENTER.
Step 7: Repeat until the five-digit password is shown on
the gauge display.
Step 8: Press ENTER.
Note: to erase password at any time while in the SETPW
(set password) mode, press the ZERO/CLEAR key. The
user will be prompted to reprogram the password once the
five-digit password is entered. The gauge will display SAVE.
Step 9: Press Enter to save the password setting.
ENTPW: Once a user password has been established and
entry into the CONFIG mode is required, the user will be
prompted to ENTPW or enter password.
Follow setup steps 4-8 above.
RECAL: or recalibrate allows for zero, mid-scale, full-scale
and factory default calibration of the gauge. The RECAL
feature also allows for recalibration of gauges with
4-20mA output.
*Indicates default.
– 9 –
(down arrow
(down arrow