Product labeling
The pressure display unit labeling, located on the back of the housing, provides key identification and
specification information (see the Figure below).
Figure 1.1: The PRT-DPM-3T pressure display unit labeling
MidJet pressure display unit Instructions for use 18117465 AD
WARNING
If you autoclave or use hot caustic cleaning solution (0.5 N NaOH) to
clean the pressure transducers, the transducers can fail causing
leaks of potentially hazardous cleaning or process fluid and bodily
harm. To avoid leaks of potentially hazardous solutions, do not
autoclave the pressure transducers. Limit exposure to caustic solu-
tion (0.5 N NaOH) to room temperature and only for short periods
(30 to 60 minutes) of time. Rinse all caustic thoroughly from the
transducers after exposure.
WARNING
If water or other liquids enter the pressure display unit, equipment
damage and electrical shock (bodily harm) can occur. Locate the
pressure display unit where it will not be exposed to water and
other fluids. Prevent water or other liquids from entering the unit.
1 Introduction