- ページ 12

計測機器 EG&G ORTEC 463のPDF 操作・サービスマニュアルをオンラインで閲覧またはダウンロードできます。EG&G ORTEC 463 20 ページ。 Constant fraction discriminator

rise time of <2 nsec and an amplitude of 10 V, the 463
walk should be tested for the range of 10 V to 100 mV.
Be careful when making this measurement to consider the
Tektronix Attenuator characteristics. It is not uncommon
for an insertion delay variation of +100 psec to exist
in the attenuator. Verify the delay consistency of the
attenuator before making the walk test, since 100 psec
is equivalent to <1/8-inch difference in signal path length
and is l ikely to exist for most attenuator units.
Na! Mode This test is identical to the procedure outlined
for Scint/SB mode except that the rise time of the signal
to the 463 should be about 5 nsec to simulate more
closely the signal from a Nal scintillator. The walk should
not exceed +0.5 nsec in this mode for a range of input pulse
heights from 10 V to 100 mV.
Walk Correction
If the unit fails to meet the walk test as
outlined above, a correction can be made. This adjustment
has been made at the factory and should not normally be
necessary unless some components have been replaced due
to failure. Use the following procedure:
1. Connect the 463 into the system shown in Fig. 6.2 and
set the Tektronix 50J2 Attenuator for XI. Move the cable
for the osci l loscope input signal to TPI on the top of the
463 and use a X10 attenuated probe.
2. Adjust R45 clockwise until the signal is similar to that
of Fig. 6.3a.
3. Adjust R45 counterclockwise until the signal is similar
to that of Fig. 6.3b.
4. Adjust R45 clockwise again until the signal at TPI is
correct as shown in Fig. 6.3c.
This provides a first-order adjustment and is usually
sufficient for proper walk characteristics. If a finer adjust
ment is necessary to minimize the walk, use the exact
system in Fig. 6.2. Set the attenuator to X50 and adjust
R45 (less than +1 turn) so that the time walk is zero.
After this adjustment the 463 should meet the walk test
suggested for al l three modes.
0 1 1 1
1 1 N
/
- /
1 1 1 1
1 1 1 1
-tHT
rt t r 11 1 1
1 1 1 1
1 1 1 1
1 M 1
1 1 1 1
(a) R45 Too Far Clockwise
(b) R45 Too Far Counterclockwise
1 1 1 1
1 1 1 r
M 41
1 1 1 1
4444-
4444
1 1 1 1
Vert.
0.5 V
/div
Horiz. 5 nsec/div
1
1
(c) Correct Adjustment
Fig. 6.3. Effect of Walk Adjustment In the ORTEC 463.