OPERATION

o
Wait for the water to heat up to 37.0
C.
For your protection, use gloves when working with any blood products.
Flat-frozen plasma units can be thawed without an overwrap bag. Hold a
bag of flat-frozen plasma (bare or overwrapped) with the ports up and the
label on the plasma unit away from the center of the CORRAL. Slip the
CLAMP (plastic bar with a stainless spring attached) over the top of the
bag and line up the 3 grooves in the CLAMP with the ports of the plasma
bag.
Place the bottom of the plasma bag into the water, between the stainless
rods of the CORRAL. Pull the stainless spring of the CLAMP over the
screw head in the CORRAL. The spring should pull the plastic bar and
plasma bag snugly against the corral. The spring tension can be adjusted
by bending the stainless steel spring. If the stainless steel spring breaks,
the clamp must be replaced (Part # C4T-CLM).
Folded frozen plasma must be overwrapped. Load it the same way. The
CLAMP will hold the plasma unit by the overwrap bag.
Push START to begin thawing for the programmed time. The Rocking
Motor will turn on. The water in the bath will be forced to flow through
the large holes in the dividing plate of the CORRAL and will massage the
plasma bags.. Use PLUS or MINUS to increase or decrease the thawing
time. When the time is finished, the buzzer will call the operator back.
The STOP switch ends the thawing. If more thawing time is needed, hold
PLUS while pushing START and the unit will thaw for another 3 minutes.

LEAKY PLASMA BAG

If a Plasma bag leaks, turn the unit off and discard the leaking bag. The
remaining bags are not contaminated, because the ports are always kept
out of the water. Dry off the plasma bags with paper towels and wipe with a bleach towelette or a 10% bleach
solution. The bath water is contaminated and must be drained and the bath cleaned. See MAINTENANCE.