Ambu 480500000 사용 설명서 지침 - 페이지 9
{카테고리_이름} Ambu 480500000에 대한 사용 설명서 지침을 온라인으로 검색하거나 PDF를 다운로드하세요. Ambu 480500000 10 페이지. Ascope 5 broncho sampler set (part 2 of 2)
6. Troubleshooting
If problems occur with the system, please use this troubleshooting guide to identify the cause
and correct the error�
Problem
Absent or reduced
suction or sampling
capability�
Unable to connect
sample container,
sampler bridge, and/
or endoscope�
Tubing cannot
be attached to
Sampler bridge�
Possible cause
The sample container is not
attached, or poor connection
of the sample container�
Sampler bridge may not be
connected properly to
the aScope™ 5 Broncho�
The suction pump is not turned
on or not connected�
Vacuum suction tubing
is not connected properly to
the suction connector or
the suction adapter�
The rubber seal may have come
off the connection interface�
The attachment lock may be
closed before attaching the
sampler bridge to an endoscope�
The sample container lid may be
closed before insertion�
Incorrect configuration of tubing
(male to the male suction
connector)�
Recommended action
Connect or re-connect the
sample container to the
sampler bridge�
Re-connect the sampler bridge
to aScope™ 5 Broncho and push
in the attachment lock to secure
the connection�
Turn the pump on and check
the suction line connection�
Ensure that the parts are
properly connected� If suction
is still absent or reduced
change to another vacuum
suction tubing�
Re-insert the rubber seal
manually� However, if it is
believed that the rubber seal
may be in contact with an
unclean surface, it may be
necessary to replace the
sampler bridge�
Pull out the attachment lock
before attaching sampler
bridge to endoscope�
Replace the sample container
with the spare provided�
Do not attempt to open the
sample container lid once closed�
The suction adapter can be
used to allow male-to-male
connections�
9