Ambu aScope 사용 설명서 지침 - 페이지 13

{카테고리_이름} Ambu aScope에 대한 사용 설명서 지침을 온라인으로 검색하거나 PDF를 다운로드하세요. Ambu aScope 17 페이지.

Ambu aScope 사용 설명서 지침
8.2.2 Advisory screen symbols on the Ambu® aScope™ Monitor
Total lifetime of the Ambu® aScope™
Total operating time of the Ambu® aScope™
Battery symbol indicator on the Ambu® aScope™ Monitor
Start-up picture
Scope time expired

9. Connecting the system

The connection between the Ambu® aScope™ and the Ambu® aScope™ Monitor is illustrated below.
1. Connect the plug on the Ambu® aScope™ cable to the socket on the front of the Ambu® aScope™ Monitor.
2. Push the ON/OFF button on the Ambu® aScope™ Monitor to start up the system. After about 10 seconds the Ambu® aScope™
Monitor is ready to display the view.
3. Push the ON/OFF button on the Ambu® aScope™
Figure 3 - Connecting the Ambu® aScope™ and the Ambu® aScope™ Monitor
8 hours from first switching on.
30 minutes from first switching on. Pay attention to the red symbol
(less than 5 minutes left) on the Ambu® aScope™ Monitor.
Recharge the Ambu® aScope™ Monitor when the green battery symbol
changes to orange
Please wait...
Ambu® aScope™
operation time icons
Full time left - 8 hours. Operation
time 30 minutes (green).
Half time left (orange)
Time is expiring. Less than
5 minutes left (red)
Ambu® aScope™
Monitor battery icons
Fully charged or above half
battery capacity (green)
Half battery capacity
(orange)
Low battery capacity (red)
Battery error (the power
plug is connected but the
battery is not charging)
Plug icons
Plug is connected and the
monitor battery is above half
charged (green)
Plug is connected and the
monitor battery is half charged
(orange)
Plug is connected and the moni-
tor battery power is low (red)
13