3. Hold the turbine flowmeter body with the rear (outlet) end down, and slide the
rotor assembly and rear support assembly into the turbine flowmeter body.
Make sure the vane is placed between the notches.
4. Once the rear support assembly is in place, install the retaining ring.
5. Blow into the turbine flowmeter to ensure the rotor spins freely.
NOTE: Installation of the replacement kit is complete and the turbine flowmeter
can be reinstalled into the process line.
6. Install the magnetic pickup. HAND TIGHTEN ONLY.
NOTE: Be sure to use the proper magnetic pickup, cable and connector for the
application.
7. Enter the new K-factor supplied with the replacement kit into the electronic
readout device. (Refer to the calibration report.)
NOTE: All internal replacement kits are factory calibrated and are supplied with a
five point calibration certificate.
Single Post Design Replacement Kit for FLOMEC G-Series Sanitary Turbines
Figure 2