1. To test if the monitor is operating correctly, you can do a simple test. Normal
Indoor O2 Levels is about 21%.
2. To test the device, simply apply pure Nitrogen on the hole located on the top
right corner of the SEU.
3. *Verify that the alarms are sounding, and the lights are flashing, on both the
SEU and RDU.
4. Now you know the monitor is working, the alarm will stop once the O2 returns
to a normal level.
(You can also press the reset alarm button on the SEU to silence the alarm.)