Display
Icons and Symbols
Icon/Symbol
20.9%
DIAG
AL1
AL2
AL3
CALI
RCFS
HI
6 | P a g e
R A D - 0 0 0 2 - Z R / M a n u a l a n d I n s t a l l a t i o n
G u i d e
O2 Level Icon: Displays the live ambient oxygen levels of the area being
monitored. Updates every 2 seconds.
Technician diagnostics feature. Communication test.
st
1
Alarm Icon: Configurable 1
2
nd
Alarm Icon: Configurable 2
3
rd
Alarm Icon: Fixed 3
rd
Calibration Icon: (See page 13 for calibration instructions)
Reset Factory Setting Icon: (See page 14 for reset instructions)
High O2 Icon: Indicated O2 levels greater than 23%
Alarm Icon: Appears after AL1 is triggered and will remain on display
while the monitor is in alarming state
Ventilation Icon: O2 Levels drops lower than AL2 and Relay2 has been
triggered.
Description
st
alarm. (19.5% O2)
nd
alarm. (16.5% O2)
alarm. (Default 12% O2)