1.
2.
Operator's Manual
From the CoaguChek XS PT Controls
Carefully remove the screw cap
Kit (stored in the fridge), take out a
and rubber stopper from the quality
bottle of control containing the dried
control bottle and make sure that
powder and a diluent filled dropper.
none of the dried powder is removed
Check the expiry date.
with the rubber stopper.
7.
8.
Touch the Scan button and hold the
Press Control Test on the main
operator ID barcode approximately
menu screen.
10cm from the built-in scanner. The
meter beeps once the barcode has
been successfully read.
0508 MY LIFE | 0508 69 54 33
For Assistance Phone
Running Liquid QC - Quick Reference Guide
3.
4.
Holding the dropper at the bottom of
the stem use scissors to cut off the
tip of the dropper at the upper end of
the stem. Do not squeeze the bulb of
the dropper.
9.
10.
Select the corresponding code for the
QC test displayed on the screen or a
new code if using a new lot of control
and insert quality control code chip
when prompted.
The meter stores the code chip
information in the memory for future
use.
Protection against infections: There is a potential risk of infection. Medical staff and other persons using the CoaguChek XS Pro
instrument to perform measurements for more than one patient must be aware that any object coming into contact with human blood is a
potential source of infection.
5.
Insert the dropper tip first, into
Gently swirl 2-3 times.
the control solution bottle. Gently
squeeze the bulb to dispense the
Minimum reconstitution time one
entire contents of the dropper onto
minute. Use within 30 minutes of
the dried powder. Do not touch the
making up.
dried material and the dropper.
11.
The dropper icon flashes to indicate
Mix the control again by gently
that the meter is ready to perform the
swirling. Draw the solution into the
test.
dropper and apply one hanging drop
to the clear target area and the meter
At the same time a 180 second
will beep when the control sample is
countdown begins. You must apply
detected and the INR measurement
the sample within this time.
begins.
6.
Slide the test strip into the test strip
guide in the direction of the arrows
until it stops.
The meter turns on and the main
menu appears.
If you insert a strip from a new lot
number for the first time, the meter
will display the number of the test
strip code chip required. Remove
the old code chip and insert the new
one.
12.
The result will appear and the QC
range will be underneath in brackets.
If this is out of range, indicated by
an up or down arrow, repeat the test
mixing a fresh vial of QC material. if
still out of range contact 0508 MY
LIFE | 0508 69 54 33.