3. Set the Language
Language
09:15 am
Dansk
Deutsch
English
Español
Français
03/16/2010
ENGLISH is the default selection. It is
highlighted in black.
Press
to save your language setting.
The display changes to the Date and Time
Format screen.
PREPARING FOR A TEST
1. Gather the Following Items
• CoaguChek XS Pro Meter
• Container of test strips
• Test Strip Code Chip
• CoaguChek Lancet
• Alcohol wipe
• Cotton ball
TESTING
1. Remove a Test Strip
Take a test strip out of the container.
Close the container tightly.
You have 10 minutes to use a test strip
once you remove it from the container.
6. Record Result
Record your result as instructed by your
facility.
Note: If during testing the meter
displays an error message, refer to the
Error Messages section of the user
manual for an explanation and steps on
how to proceed.
CCXS_Pro_GSG_05914647001_03_USA.indd 1
XS Pro
®
Getting Started
4. Set the Date and Time Format
5. Set the Date
Format
09:15 am
Date:
DD.MM.YYYY
Year:
MM/DD/YYYY
Month:
YYYY-MM-DD
24H
12H
Time:
Day:
03/16/2010
03/16/2010
MM/DD/YYYY is the default date
Press
selection.
12H is the default time selection
Press
to save your date and time format
settings.
The display changes to the Date screen.
Press
The display changes to the time screen.
The date settings are important. Each time
you run a test, the meter compares its date
with the test strip's expiration date. An
incorrect date may cause an expired strip
error on in-date strips preventing you from
running a test.
If the test strips are expired, the meter
displays an error message and prevents you
from running a test.
2. Match Code
3. Insert Test Strip Code Chip
The code number on the test strip
With the code number facing up, insert
container and the test strip code chip
the code chip into the code chip slot
must match.
until it snaps into place.
Each box of test strips comes with a
matching code chip. Test strip code
chips are identified with an "S."
Every time you open a new box of test
strips, you will replace the test strip
code chip.
2. Insert Test Strip
3. Select PATIENT TEST
Slide the test strip into the test strip
Press PATIENT TEST on the main menu
guide in the direction of the arrows until
screen.
it stops.
Enter the Patient ID.
The meter powers on, and the main
An hourglass symbol appears.
menu appears.
A test strip symbol appears with a
If you insert a strip from a new lot
flashing blood drop. You have 180
number for the first time, the meter
seconds to apply the sample.
will display the number of the test strip
code chip required. Remove the old
code chip and insert the new one.
7. Clean Up
Place the used strip and lancet in an
approved biohazard container.
Power the meter off.
If the meter is dirty, wipe it clean.
GETTING STARTED
Follow these steps to get started using the meter:
1. Watch the CoaguChek XS Pro System DVD. It will help you get comfortable with the
CoaguChek XS Pro Meter and the testing procedures.
2. Read this Getting Started guide. You will learn how to set up and perform your first blood
and optional liquid quality control tests on the CoaguChek XS Pro Meter.
3. With this Getting Started guide by the meter, follow the steps to perform your first test.
Then, as necessary, refer to the user manual:
The CoaguChek XS Pro System User Manual is a comprehensive guide to the meter and
test strips. It is designed to provide answers to your questions about the meter's operation
and use.
INSTALLING BATTERIES
1. Open Compartment
Press the release tab on the battery
cover on the back of the meter.
Slide the battery cover off.
After you insert the batteries, the meter powers on and the display changes to the
language menu.
You'll need to set several parameters before using the meter for the first time.
6. Set the Time
Date
09:15 am
Time
2010
Hour:
1
Minute:
1
03/16/2010
or
to select the year, then
Press
or
to select the month, then
or
to select the day.
then AM or PM.
to save the current date.
Press
The display returns to the Main Menu.
4. Wash Hands
Have the patient wash his or her hands
in warm, soapy water. Or, clean the
fingertip with an alcohol wipe.
Make sure the fingertip and your
gloved hands are thoroughly dry.
4. Collect the Sample
Twist the protective cap off the lancet.
Massage the finger until you see
increased color in the fingertip.
Keeping the hand down, press the
tip of the lancet firmly against the
side of the fingertip. Press the blue
trigger button.
Gently squeeze from the base of the
finger to develop a drop of blood.
2. Insert Batteries
Insert 4 AA batteries according to
the diagram inside the battery cover.
Replace the cover.
MAIN MENU
PATIENT TEST
QC TEST
MEMORY
MAIN MENU
Screen
Contrast
SETUP
09:15 am
PATIENT TEST
9
Setup
Setup
Units
59
08:00 AM
QC TEST
Screen
Screen
Language
MEMORY
Screen
Options
AM
PM
Options
Contrast
SETUP
Date/Time
ID Setup
ID Setup
Units
You may change or re-enter the date,
08:00 AM
08:00 AM
or
to select the hour, then
Lockouts
Lockouts
time or format at any time. Always access these
Language
options by:
or
to select the minutes,
08:00 AM
Pressing Setup on the main menu, then
08:00 AM
Date/Time
Screen on the setup menu, then
to save the current time.
Date/Time on the Screen menu
08:00 AM
...which brings you to the SET DATE,
SET TIME, and FORMAT options.
Before continuing, review these
tips for getting a good blood drop.
Increasing the blood flow in the finger
will help you get a good drop of blood,
so keep in mind these tips:
• Warm the hand. Have the patient hold
it under his or her arm, use a hand
warmer, and/or wash with warm water.
• Have the patient let that arm hang by
his or her side.
• Massage the finger from its base.
Use these techniques until the fingertip
has increased color.
5. Apply the Sample
Within 15 seconds of sticking the
fingertip, apply the blood to the
target area on the test strip.
Find the clear target area on the test strip.
You can dose from the side or top.
Hold the blood drop to the test strip
until you hear a beep. The flashing
blood drop symbol will disappear and an
hourglass symbol appears and the test
begins.
Do not add more blood to the test
strip. Do not touch the test strip
while the test is in progress.
The meter performs an automatic quality
control check on the strip before it
displays the test result.
QC
appears in the display briefly
following a successful QC check.
The result appears in less than a minute.