b. Press the PLAY/PAUSE button again to set the treatment time.
c. Press the PLAY/PAUSE button again to start the treatment.
EXAMPLE:
1. An example time of one minute and twenty-three seconds will be used (01:23).
2. The right most character should be flashing – press the UP-ARROW button three (3)
times. The display should now read "00:03".
3. Press the LEFT ARROW button one (1) time and the UP-ARROW button two (2) times.
The display should now read "00:23".
4. Press the LEFT ARROW button one (1) more time and the UP-ARROW button one (1)
more time. The display should now read "01:23".
5. Press the PLAY/PAUSE button one time to lock in the treatment time. The flashing
character should stop flashing.
6. Now that the desired time is displayed (in this case 01:23), press the PLAY/PAUSE button
to begin treatment.
7. Pressing the PLAY/PAUSE button again will pause treatment.
8. While paused, pressing the CANCEL button will end the treatment and prompt a new
treatment.

12.0 Viewing Device Code for Flex Rx

The Device Code is predetermined by the software. To view the next code on the Digital Timer
controller:
1. Enter your Key Code to unlock the controller.
2. Simultaneously press Cancel, left, and play/pause.
3. While the screen displays "- - - 1", press the UP-ARROW button four (4) times. The screen
should display "- - - 5".
4. Press the PLAY/PAUSE button to enter Menu 5.
5. The value displayed ("Cxyz") is the next generated Device Code.
a. Alternately, when all treatments have been exhausted, the display will display this
Device Code.
6. If all treatments have been exhausted and a wrong refill code is entered, the Digital Timer
Controller will flash the entered code two (2) times then display the Device Code.

13.0 Enable/Disable/Refill FlexRX

The Flex Rx exposure limiting system uses a "call and response" method. The 3-digit Device
Code (Cxyz) is randomly generated by the controller and is necessary to generate a 4-digit
Treatment Code using an algorithm controlled by Daavlin.
MNL-00049 [1]