Dongguan Xin Bao Instrument SBT401 운영 매뉴얼 - 페이지 16

{카테고리_이름} Dongguan Xin Bao Instrument SBT401에 대한 운영 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. Dongguan Xin Bao Instrument SBT401 16 페이지. Abrasion and pilling tester

Dongguan Xin Bao Instrument Co., Ltd
set of photographs is also available from SBT which are mounted in an assessment viewer to enable
the specimens to be more consistently rated.

SECTION 6 CALIBRATION AND MAINTENANCE

6.1 Calibration

1.
The motion plate and abradant plate should be parallel. Insert a dial gauge through the sample
holder bearing and inch the top plate.
The needle movement over the dial gauge should be within ±0.05mm over the whole surfaces
of the abradant tables (with the abradant cloth removed).
2.
Assemble each sample holder complete with the weight and place each one in position on the
appropriate table.
Use a feeler gauge to check for any gap between the face of the specimen holder and the table.
3.
to adjust tables, slacken the central screw and adjust as appropriate using the three jacking
screws located on the main deck plate.
When complete, re tighten the central securing screw.
4.
to adjust the motion plate, slacken the locking nut fitted to the stem of the three ball carrying
cups rotate the cups to the correct height and re tighten the locking nuts.

6.2 Maintenance

1.
The motion plate slides should be lightly smeared with grease.
2.
The ball cups should be lightly oiled as required.
Add: The 3rd Industrial Zone, Chashan Town, Dongguan. Tel: 0769-33365786 www.ots-testequipment.net