AmeriWater 00HC-0091 운영 및 유지보수 매뉴얼 - 페이지 3

{카테고리_이름} AmeriWater 00HC-0091에 대한 운영 및 유지보수 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. AmeriWater 00HC-0091 10 페이지. Healthcare storage tanks

1.0

STORAGE TANK COMPONENTS

THEORY OF OPERATION
The storage tank is equipped with 3 float switches, which will turn the
reverse osmosis unit off when the water reaches the predetermined
cutoff point and turn the RO on when the tank is drained to the
appropriate level float. The distribution pump on the storage tank will
cycle whenever there is a pressure drop detected. A third float
switch has been wired into the pump to act as a kill switch to prevent
the pump from operating when there is no water. (Applies to
models, 00HC-0091, 00HC-0093, 00HC-0095 and 00HC-0097)
The storage tank is equipped with 2 float switches, which will turn the
reverse osmosis unit off when the water reaches the predetermined
cutoff point and turn the RO on when the tank is drained to the
appropriate level float. The distribution pump on the storage tank will
cycle whenever there is a pressure drop detected. (Applies to
models, 00HC-0090, 00HC-0092 and 00HC-0094)
The storage tank is equipped with 2 float switches, which will turn the
reverse osmosis unit off when the water reaches the predetermined
cutoff point and turn the RO on when the tank is drained to the
appropriate level float. The distribution pump on the storage tank will
cycle whenever there is a pressure drop detected. (Applies to
models, 0185-8000 and 00HC-0090-PRV)
The Healthcare storage tanks come in three different sizes – 55
Gallon, 100 Gallon and 185 Gallon. Shown in this manual is the 185
Gallon model.
1