1.0
STORAGE TANK COMPONENTS
THEORY OF OPERATION
The storage tank is equipped with 3 float switches, which will turn the
reverse osmosis unit off when the water reaches the predetermined
cutoff point and turn the RO on when the tank is drained to the
appropriate level float. The distribution pump on the storage tank will
cycle whenever there is a pressure drop detected. A third float
switch has been wired into the pump to act as a kill switch to prevent
the pump from operating when there is no water. (Applies to
models, 00HC-0091, 00HC-0093, 00HC-0095 and 00HC-0097)
The storage tank is equipped with 2 float switches, which will turn the
reverse osmosis unit off when the water reaches the predetermined
cutoff point and turn the RO on when the tank is drained to the
appropriate level float. The distribution pump on the storage tank will
cycle whenever there is a pressure drop detected. (Applies to
models, 00HC-0090, 00HC-0092 and 00HC-0094)
The storage tank is equipped with 2 float switches, which will turn the
reverse osmosis unit off when the water reaches the predetermined
cutoff point and turn the RO on when the tank is drained to the
appropriate level float. The distribution pump on the storage tank will
cycle whenever there is a pressure drop detected. (Applies to
models, 0185-8000 and 00HC-0090-PRV)
The Healthcare storage tanks come in three different sizes – 55
Gallon, 100 Gallon and 185 Gallon. Shown in this manual is the 185
Gallon model.