Pump draws excessively high current 
 Motor is overloaded 
 Power input is incorrect 
Low or zero flow 
Pump is rattling or knocking 
Servicing 
Listed  below  are  the  two  predominant  types  of  maintenance  typical  for 
servicing of the consumable parts (diaphragm, valve discs, gasket, and Teflon® washer), and the servicing 
of  the  connecting  rod.
http://www.airdimensions.com/service/videos/.
Disassembly of Head Section and Diaphragm 
Remove head assembly by unscrewing the four head bolts. A flat‐bladed screw driver may be needed to 
gently pry the head free of the service diaphragm. **If the heads have Teflon® coating use caution not to 
scratch the surface. 
 
The valve body can then be removed by unscrewing the two smaller screws (also accessible on the top of 
the head section). This part may be freed by gently tapping on these two screws after they have been 
loosened about three or four turns. When the valve body is removed, check all internal surfaces for any 
accumulation  of  dirt.  The  two  valve  discs  can  be  wiped  clean  and  replaced  as  long  as  they  appear 
unaffected by usage. The valve gasket can be easily removed and should be inspected. As a matter of good 
 
 Turn off pump 
 Remove all pressure and vacuum conditions 
 Restart and test at atmospheric pressure 
 Check motor wiring i.e. 115 V vs 230 V wiring 
 Check power source 
o Pumps are only rated for ±10% voltage on name plate 
Connection or sample line blocked 
Remove blockage 
Liquid or foreign debris has collected in the head 
Clean out the head, see section "Servicing" 
Place pump outlet facing downward 
Diaphragms, valves, or gasket are worn 
See section "Servicing" 
Connecting rod bolt or diaphragm plate screw are under torqued
See section "Servicing" for torquing specifications
Connecting rod cap is too close to one side of housing
Using a screw driver lightly pry the cap away from the side of the housing 
and center. A centering tool is available for purchase at ADI. 
 
For  video  instructions  on  servicing  the  head  and  diaphragm  visit 
 
Dia‐Vac®  Pumps,  the