Ames Instruments 64020 사용 설명서 및 안전 지침 - 페이지 11

{카테고리_이름} Ames Instruments 64020에 대한 사용 설명서 및 안전 지침을 온라인으로 검색하거나 PDF를 다운로드하세요. Ames Instruments 64020 16 페이지. 400a ac digital clamp meter

Continuity Measurement
Test continuity between two
points of a circuit.
WARNING! To prevent electric
shock, disconnect all power to
circuit and discharge all high-voltage
capacitors before measuring.
1.
Plug black Test Lead into
black COM Jack.
Plug red Test Lead into red Input Jack.
2.
Turn Rotary Switch to
Capacitance Measurement
Measure capacitance up to 100μF.
WARNING! To prevent electric
shock, disconnect all power to
circuit and discharge all high-voltage
capacitors before measuring.
Note: It may take up to 30 seconds
for reading to stabilize.
1.
Plug black Test Lead into
black COM Jack.
Plug red Test Lead into red Input Jack.
2.
Turn Rotary Switch to
Item 64020
position.
position.
For technical questions, please call 1-888-866-5797.
3.
Press SELECT button
until
is displayed.
4.
Short Test Leads together, meter
should beep continuously.
5.
Connect Test Leads to two points of
circuit. If resistance value is less than
30Ω, Meter will beep continuously.
6.
When testing is complete, turn
Rotary Switch to OFF, remove
Test Leads and store with Meter.
3.
Press SELECT button
until n F appears.
4.
Press REL button to zero out
reading as necessary.
5.
Carefully touch capacitor
leads with tips of probes.
6.
Read measurement.
Note: If capacitor is short
circuited or capacitance is too
high, display will read OL.
7.
When testing is complete, turn
Rotary Switch to OFF, remove
and store capacitor and Meter.
Page 11