Ametek Jofra ASC-400 조정 매뉴얼
5.4
Adjusting current (mA) Output:
1. Connect Digital Multimeter mA input to ASC-400 mA source terminals.
2. Type "4" in the Calibration menu.
Prompt = "First calibration point. Enter mA output:"
3. Enter multimeter read value when stable (example: 0.0000) and press "enter".
Prompt = "Second calibration point. Enter mA output:"
4. Enter multimeter read value when stable (example: 24.0000) and press "enter".
ASC-400 now returns to the Calibration menu.
5.5

Adjusting Voltage input:

1. Connect Calibrator V output to Digital Multimeter V input and ASC-400 V
measure terminals.
2. Type "5" in the Calibration menu.
Prompt = "Input 0 volt to ASC400 - Press <enter> when stable"
3. Apply 0.000 V (Digital Multimeter reading) to the input and press "enter" when
stable.
Prompt = "Input 20 volt to ASC400 - Press <enter> when stable"
4. Apply 20.000 V (Digital Multimeter reading) to the input and press "enter"
when stable. ASC-400 now returns to the Calibration menu.
5.6

Adjusting Voltage Output:

1. Connect Digital Multimeter V input to ASC-400 V source terminals.
2. Type "6" in the Calibration menu.
3. Follow the menu prompts.
Prompt = "First calibration point. Enter V output:"
4. Enter multimeter read value when stable (example: 0.0000) and press "enter".
Prompt = "Second calibration point. Enter V output:"
5. Enter multimeter read value when stable (example: 20.0000) and press "enter".
ASC-400 now returns to the Calibration menu.