COM-power corporation CLCE-400 사용 설명서 - 페이지 16

{카테고리_이름} COM-power corporation CLCE-400에 대한 사용 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. COM-power corporation CLCE-400 18 페이지. Rf current probe and calibration fixture

COM-power corporation CLCE-400 사용 설명서
Page 16 of 18
INSTRUCTION MANUAL
 
When the probe is used to measure current, the transfer impedance factors
convert the [measured] voltage quantity present at the probe's output port
into a current quantity corresponding to the actual current through the
conductor(s) passing through the probe aperture.
The transfer impedance factors include the insertion loss of the probe, and
also perform the function of converting the measured voltage quantity into
a current quantity. Based on the assumption that the probe is connected
to a measuring instrument having a nominal input impedance of 50 ohms,
this conversion is simple Ohm's Law:
The transfer impedance factors are determined by combining the
conversion described above with the insertion loss calibration data
(L
), as
ins
shown below:
The transfer impedance factors can be applied in practice, as follows:
- C A L I B R A T I O N
S E C T I O N 4
Rev0305