Baumer PAC50H 운영 매뉴얼 - 페이지 13

{카테고리_이름} Baumer PAC50H에 대한 운영 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. Baumer PAC50H 44 페이지.

Baumer PAC50H 운영 매뉴얼
Baumer
6.5

Mounting the sensor

WARNING
Health hazard from contaminated media
a) Only use welding sleeves and adapters from Baumer.
b) Do not seal the process connection with Teflon tape (PTFE) or elastomer.
c) The welding should only be carried out by persons who are trained in the hygiene sector.
INFO
Weld-in sleeves or adapters in hygienic applications
It is recommended to use weld-in sleeves or adapters for sensor installation.
Observe the following:
a) 3-A mark or arrow must point upward.
b) The leakage hole must point downwards.
c) Hygienically and internally flush.
d) Grind welding to Ra ≤ 0.8.
WARNING
Danger caused by leaking system
Sensor installation in pipe systems may cause leaks. Such leaks may cause a great number of
hazards.
Therefore, it is imperative to observe the following during sensor mount:
a) Only use welding sleeves and adapters from Baumer.
b) Always check seals for damage and cleanliness.
c) The welding should only be carried out by persons who are trained in the hygiene sector.
d) Never seal process connections with Teflon tape (PTFE) or elastomer.
a) Select a suitable installation location for the sensor.
Observe the installation requirements.
b) Make sure that the pipes are free from any pressure.
c) Make sure no media can escape at the installation site.
d) Mount the sensor in the intended position.
Apply 20 - 25 Nm to screw in the sensor.
e) Check the pipe system in the sensor area upon any leak.
f) Perform electrical sensor connection.
g) Check leak-tightness of the sleeve.
h) Check M12 connector upon tightness.
V1 | PAC50H/S
Operating Manual
Installation
|
6
13