4.3

Functional description

6
Ill. 3: Functional description (schematic)
1
Oscillator
3
Output (4 - 20 mA) optionally conductivity
or temperature
5
Temperature sensor Pt100
7
Primary coil
The sensor measures the conductivity and temperature of liquids, and optional calculates con-
centration values.
Sensor operation utilizes the inductive sensing principle without mechanical or process-interfer-
ing components.
The hole in the sensor tip is surrounded by 2 coils. The primary coil is supplied with AC voltage.
The secondary coil measures the induced current of the fluid in the measuring hole. The fluid's
temperature is measured at the Pt100 sensor tip. This will ensure temperature compensation of
the conductivity signal.
4.4

Operating and display elements

System parameters and the measured values of the sensor are shown on the display.
Sensor parameterization is by the following options:
Display
n
FlexProgram
n
IO-Link master
n
Also see about this
2 Parameterization using the integrated display [} 18]
2 Parameterization via FlexProgram [} 22]
2 Parameterization via IO-Link master [} 23]
V1 | PAC50H/S
1
7
5
2
3
4
2
Select range: R1 or R2; Output selection:
conductivity or temperature (4 - 20 mA)
4
Secondary coil
6
Media
Operating Manual
Description
|
4