Baumer TER8 사용 설명서 - 페이지 5
{카테고리_이름} Baumer TER8에 대한 사용 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Baumer TER8 28 페이지. Temperature measurement, front-flush and low-invasive resistance thermometers
5.2 TER8 mounting
DANGER
Risk of injury from hazardous medium
Only use weld-in sleeves or adapters from Baumer.
f
Wear protective equipment for hazardous media (e.g. acids, alkaline solutions).
f
Do not seal the process connections with Teflon tape (PTFE) or elastomer.
f
Empty vessel and pipes before mounting.
f
Check leak-tightness of the sleeve.
f
Check tightness of the M12 plug.
f
Example of mounting with weld-in sleeve ZPW3-321
Example of mounting with weld-in sleeve ZPW2-326
Some weld-in sleeves (e.g. ZPW2-326) must be mounted in the correct angle to allow automatic draining.
170°
5°
www.baumer.com
1
1
1
2
5°
Operating Instructions
Temperature measurement
TER8
Front-flush and low-invasive resistance
thermometers
9 Hole for mounting the sensor is easily accessible
Mount weld-in sleeve or adapter as follows:
f
3-A mark or arrow points upwards
Hygienically and internally flush
Grind welding to Ra ≤ 0.8 μm.
f
Screw in sensor.
f
Tightening torque: 10 ... 15 Nm
1
ZPW3-321
2
Leakage hole
5°
190°
5°
5 / 28