Ample Scientific Champion E-33 Series 매뉴얼 - 페이지 2

{카테고리_이름} Ample Scientific Champion E-33 Series에 대한 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. Ample Scientific Champion E-33 Series 4 페이지.

Ample Scientific Champion E-33 Series 매뉴얼

4. Maintenance and Care

With proper care and maintenance, E-33 will provide years of service to
your laboratory. Please follow the recommended guidelines:
1. Always disconnect the power source before performing cleaning
and maintenance
2. Use only quality test tubes. Lower quality glass tubes are likely to
fracture, leaking its contents into the rotor buckets even under low
rpm.
3. Never force a test tube into the rotor bucket. The rotor buckets are
designed to hold most commonly sized tubes, vacutainers, and SST's.
Some tubes are too large for the rotor buckets therefore should not
be used in the E-33.
4. Keep the rotor buckets clean. If a test tube breaks inside the rotor
bucket, either dispose the entire rotor bucket and its content or
safely dispose the sample and broken glass. Make sure to
thoroughly clean and disinfect the inside and outside of the rotor
buckets.
5. It is recommended that the centrifuge to be cleaned at least once
every week. Any spilled sample in the chamber and on the
centrifuge must be immediately wiped and cleaned. Only use mild
cleaning solutions when cleaning the centrifuge. Strong and acidic
cleaners are not recommended for cleaning and should never be
used on the centrifuge. The rotor and buckets can be autoclaved up
to 120˚C for sterilization. Consult manufacturer before cleaning or
sterilizing the centrifuge with any methods not specified by the
manufacturer.
6. The centrifuge should be kept and stored in the following condition
for optimal performance:
a. Ambient temperature of 0-30˚C
b. Relative humidity of ≤ 30%
c. Atmosphere clear of electroconductive dusts, explosive gas,
and corrosives
Pg | 3

5. Prohibited Usage of Centrifuge

1. Do not store or use the centrifuge in conditions not specified
2. Do not let unqualified persons operate the centrifuge.
3. Do not operate without properly installing the rotor and rotor
buckets
4. Do not overload the rotor
5. Do not use rotor or any accessories that are damaged or corroded.
6. Do not use rotor and accessories not included in the original
package or not recommended by the manufacturer. (except for
standard test tubes)
7. Do not use chemicals that cause damage and corrosion to the rotor
and accessories.
8. Do not centrifuge medium that is flammable or explosive
9. Do not operate near explosive materials
Note:
1
– Incorrect power input will damage the centrifuge, preventing proper
operation
2
– Overloading the centrifuge will prevent motor from working properly and
shortening centrifuge life.
3
– Putting the centrifuge near explosives increases the risk of centrifuge
exploding
2
3
1
.
3
Pg | 4