Doza DROZD 사용자 설명서 - 페이지 8
{카테고리_이름} Doza DROZD에 대한 사용자 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Doza DROZD 11 페이지. Dosimeter
4 CALIBRATION ROUTINE
4.1 General requirements
4.1.1 Calibration of a dosimeter is performed in accordance with the IEC 61453:2007.
4.2 Preliminary arrangements
4.2.1 Operations performed during calibration are listed in the Table 4.1.
Table 4.1 – List of calibration operations
1 External examination
2 Testing
3 Determination of basic relative measurement error of AEDR of gamma
radiation
4 Determination of basic relative measurement error of AED of gamma
radiation
4.2.2 Primary and auxiliary tools and equipment necessary for calibration are presented in
the Table 4.2.
Table 4.2 – Primary and auxiliary tools and equipment for calibration
4.5.3, 4.5.4
Calibration installation UPGD-2M-D or equivalent, with
source; range of reproduced AEDR – from 1·10
not more than ±5 %.
4.5.3, 4.5.4
Stopwatch with measurement range 1 to 3600 seconds
4.5.3, 4.5.4
Thermometer with measurement range 0 to 30 °C and scale interval 0.1 °C
4.5.3, 4.5.4
Psychrometer with measurement range 20 to 90 %, error ±6 %
4.5.3, 4.5.4
Barometer with measurement range 60 to 120 kPa and scale interval 1 kPa
N o t e - It is allowed to use other tools and equipment with similar characteristics ensuring determination of
metrological characteristics of dosimeters with required precision.
4.3 Safety precautions
It is necessary to follow safety requirements described in section 3.2 and in documentation
accompanying calibration tools and equipment.
4.4 Conditions
The following normal operating conditions shall be met during calibration:
-
air temperature ............................................................................ +(20 ±5) °C;
-
relative air humidity .................................................................. from 30 to 80 %;
-
atmospheric pressure .............................................................. 86.0 to 106.7 kPa;
-
natural radiation background ............................................ not more than 0.2 μSv·h
Operation
Name and type (designation) of primary or auxiliary calibration tool, document
that regulates technical requirements and/or metrological and main technical
FVKM.412113.026RE
characteristics of the calibration tool
8
Section
4.5.1
4.5.2
4.5.3
4.5.4
137
Cs source, with
-5
-2
-1
to 5·10
Sv·h
with uncertainty
137
Cs
-1
.