Daihan Scientific UniFreez U25 사용자 설명서 - 페이지 14

{카테고리_이름} Daihan Scientific UniFreez U25에 대한 사용자 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Daihan Scientific UniFreez U25 20 페이지. Personal digital ultra low temperature freezer

Daihan Scientific UniFreez U25 사용자 설명서
DAIHAN-Sci.com or
ALL
l Controller
Controller Display
< Fig.3.2.5>
DAIHAN Scientific, Ultra Low Temperature Smart Freezer
for
LAB.com
Shopping Mall
Controller Display
PV : Present Value of temperature inside the freezer
SV : Set Value of temperature required by users
OUT 1 : Illuminated while a compressor is operating
OUT 2: Illuminated when a over temperature alarm is activated
AL 1, 2 : Illuminated when the event output is " ON"
COM : Status of RS232/C communication, Illuminated in
available status
(RS232/C communication is served as optional request)
RDY : Ready to operate, Illuminated in operative status
DI : Illuminated in "DI ON"
Auto/Manu Button : Use for saving certain value and going
into "controller configuration mode"
SET Button : Use for setting temperature value
Button : Increase and decrease set temperature value.
▲ ▼
* Notice : Please don't adjust any value in "Parameter
Set", if you adjust it, it may Cause bad performance or
defect of unit
NO
Power Failure Function Switch
Over Temperature Function
Switch and Warning Lamp
LCO2 Inflow Function Switch
<Fig.3.2.4> Controller for UniFreez
Names
Main Controller
LCO2 Test Switch
and Warning Lamp
and Lamp
TM
U25
14/40