35XP-A Electrical

Specifications

(at 23°C ± 5°C, < 75 % R.H.)
DC VOLTS
Ranges: 400 mV, 4 V, 40 V, 400 V,
1000 V
Accuracy: ± (0.5 % rdg + 1 dgt)
AC VOLTS (45 Hz to 500 Hz)
Ranges:400 mV, 4 V, 40 V, 400 V,
750 V
Accuracy: ± (1.5 % rdg + 5 dgts) 45 Hz
to 100 Hz on 400 mV range: ± (1.5 %
rdg + 5 dgts) on 4 V to 750 V ranges
DC CURRENT
Ranges: 400 µA, 4000 µA, 40 mA,
400 mA, 2 A
Accuracy: ± (1.0 % rdg + 2 dgts) on
400 µA to 400 mA ranges: ± (2.0 % rdg
+ 3 dgts) on 2 A range
AC CURRENT (45 Hz to 500 Hz)
Ranges: 400 µA, 4000 µA, 40 mA,
2A
400 mA,
Accuracy: ± (1.5 % rdg + 5 dgts) on
400 µA to 400 mA ranges: ± (2.5 % rdg
+ 5 dgts) on 2 A range
RESISTANCE
Ranges: 400 Ω, 4 kΩ, 40 kΩ, 400 kΩ,
4 MΩ, 40 MΩ
Accuracy: ± (1.0 % rdg + 4 dgts) on
400 to 4 MΩ ranges: ± (3.0 % rdg +
5 dgts) on 40 MΩ range
CONTINUITY
Audible indication: Less than 25 Ω
DIODE TEST
Test current: 1.2 mA (approximate)
Accuracy: ±(1.5 % rdg + 3 dgts)
Open circuit volts: 3.0 dc typical
CAPACITANCE
Ranges: 4 nF, 40 nF, 400 nF, 4 µF,
40 µF, 400 µF, 4 mF
Accuracy: ± (5.0 % rdg + 30 dgts) on
4 nF range: ± (5.0 % rdg + 5 dgts) on
40 nF and 400 µF ranges: ± (5.0 % rdg
+ 15 dgts) on 4 mF range
TEMPERATURE
Ranges: -20°C to 1000°C, -4°F to
1832°F
Accuracy: ± (2.0 % rdg +4°C) -20°C to
10°C
± (1.0 % rdg + 3°C)10°C to 200°C
± (3.0 % rdg + 2°C) 200°C to 1000°C
± (2.0 % rdg + 8°F) -4°F to 50°F
± (1.0 % rdg + 6°F) 50°F to 400°F
± (3.0 % rdg + 4°F) 400°F to1832°F
FREQUENCY
Ranges: 4 k, 40 k, 400 k, 1 MHz
Accuracy: ± (0.1 % rdg + 3 dgts)
Sensitivity:
10 Hz to 1 MHz: >2.5 V rms
OVERLOAD PROTECTION
Voltage, Resistance, Diode, Continuity,
Frequency, Temperature: 1000 VDC or
750 VAC rms
Current, Capacitance: 2 A / 1000 V fast
blow ceramic fuse 6.3 × 32 mm
mA JACK: Input warning detects wrong
switch/input jack configuration
AUXILIARY FEATURES
DATA HOLD: Freeze the latest reading
on the display.
RANGE: Execute manual range mode.
AUTO Power off: After auto power off,
press (RANGE) button to restart the
meter, and the last reading of
measurement will be returned to the
display.
12