Auria Processor Troubleshooting
Determining Device Function
System Sensor
The system sensor, a handheld device that performs a system functionality test, is
for use with the Auria, Platinum BTE/cII BTE, PsP, and s-series sound Processors.
The system sensor is powered by a lithium battery, which needs to be replaced
approximately every two years.
To Use the System Sensor:
Place the system sensor over the headpiece at a distance of approximately one (1) to
two (2) inches, then gently press the button. The system sensor will display its status.
•
OrAnGE
light indicates that the system sensor itself is functioning properly.
•
No OrAnGE
light indicates that the system sensor battery is depleted. You will
need to obtain a new system sensor.
once you have verified the system sensor status (orange light), slowly move the
sensor toward the headpiece and observe the system-transmitting status as follows:
•
Solid GrEEn
light means that the system is functioning properly.
•
Flashing GrEEn
light means that the Processor is not communicating with the
implant. The Processor battery is charged, but the system is not successfully
transmitting a signal to the Implant. stop and proceed to Troubleshooting
section of this guide.
•
No GrEEn
light indicates that the Processor battery is discharged and needs to
be changed.
Tools for schools by Advanced Bionics
9
AURIA® soUnd PRocEssoR Product Guide
system sensor