DPM TT 570 A 운영 매뉴얼 - 페이지 5

{카테고리_이름} DPM TT 570 A에 대한 운영 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. DPM TT 570 A 16 페이지. Micromanometer

 
On / Off: 
 
Press and hold ENABLE then press ON or OFF as appropriate.
Auto Zero: 
 
The cycle will begin as soon as the instrument is switched on.  This is factory set at a 30
second interval upon switch on and thereafter at 60 second intervals.   
To change this timing see User Menu page 05. 
The instrument contains a miniature solenoid valve, which isolates the pressure sensors 
during auto zero. The valve emits two clearly audible clicks, which signal the start and end of 
the auto zero process. 
Battery Low: 
 
The battery condition is monitored every time the instrument auto zeroes. 
When the 'battery low' warning appears, the battery must be replaced immediately, 
otherwise the readings obtained will be unreliable.  
If at any time the readings seem suspect, check the condition of the battery by zeroing the 
instrument manually using ZERO.   
If in doubt replace the battery.  For rechargeable instruments charge the internal battery 
with the charger supplied. 
Units: 
 
The unit of measurement is shown on the right hand side of the display. 
Being an auto ranging instrument, the correct resolution and the decimal point will be 
displayed according to the pressure being applied. 
FAST SLOW: 
 
Some pressurised circuits exhibit pressure and velocity fluctuations, which can be 
disconcerting; if unacceptable fluctuations occur, press FAST SLOW and verify that the slow 
indication is displayed on the screen. 
Store: 
 
Shows the number of readings stored on the instrument.
To store data manually press STORE while the instrument is in Slow1, Slow2, Slow3 or Slow4 
mode: the reading on the display will be logged and the counter will increase by one.  
During data logging the counter will increase by one every time the instrument records a 
value.  For information on data logging see Option Menu pages 08 and 09. 
Audible Feedback: 
 
While the instrument is on, audible feedback can be heard every time a key is pressed.
Signal In / Reference: 
 
Pressure ports. The instrument responds to positive, negative and differential pressure.
RS 232 Port: 
 
Is located between the two ports.
 
MODE OF OPERATION