1.3

Specimen Stage

The stage has an adjustable height limit. This is used to prevent the stage from being raised too high which could result
in damaging the specimen or even the objective lenses. The adjustment screw is located on the body just behind the
stage.
Rotate the objective turret until the 40X objective lens is in place. With a specimen mounted on the stage, slowly raise
the stage using the coarse-focus knob until the specimen is a few millimeters from the lens. Use the fine-focus knob to
continue raising the stage until the specimen barely comes in contact with the lens. The height-adjustment screw should
be tightened to set the stopping position. If the specimen was too far from the lens, and could not make contact, loosen
the adjustment screw slightly, then continue to make fine focus adjustments until the specimen is high enough. Once
the height is properly set, tighten the adjustment screw.