AmScope SE305R-P 사용자 설명서 - 페이지 18
{카테고리_이름} AmScope SE305R-P에 대한 사용자 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. AmScope SE305R-P 19 페이지.
SE300 Series
Common Issues (Optical)
Symptom
Obstructions are
Stains, dust, or dirt has accumulated on the speci-
men
observed in the field
of view
Stains, dust, or dirt have accumulated on the
lenses
Stain or dust has accumulated on the lens in the
Unclear Image
inlet of the head
The microscope head is not in the right position
to be focused
One side of the field
The specimen is not fixed
of view is dark or the
image moves while
focusing
The light source is not present
The field of view is not
bright enough
The light source is too dim
Stains, dust, or dirt has accumulated on the con-
denser, objective, eyepieces, or base lens
Cause
OPTICAL ISSUES
18
Troubleshooting
Solution
Clean the slide or use a new specimen if sample is
destroyed
Clean the objective and eyepiece lenses
Clean the lens with lens cleaner or a nonabrasive
lint free cloth, as well as spray with compressed air
Adjust the height of the microscope with the focus-
ing knob on the focusing rack until image appears
in focus. Note: Higher magnification will require
the head to move closer, while lower will require
more distance between the sample and the lens.
Secure the slide to the stage with clips or another
manner so that the sample does not move
A light source must be used with this model micro-
scope to view a sample (ring light, built in illumina-
tor, or gooseneck illuminator)
The light source used is not bright enough, or is set
to too low of an intensity setting. Adjust setting to a
higher or change light sources
Thoroughly clean tall lenses with lens cleaner or
a lint free nonabrasive cloth