Amsino AMSure SENTINEL ENF500K 매뉴얼 - 페이지 16

{카테고리_이름} Amsino AMSure SENTINEL ENF500K에 대한 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. Amsino AMSure SENTINEL ENF500K 16 페이지. Enteral feeding pump

Sentinel® Enteral Feeding Pump Quick Reference
Refer to Operating Manual for detailed instructions
See also Programming Directions on the pump
Reset Volume Delivered
(current dose) to Zero
1
Hold for 3 beeps
2
3
Reset Accumulated
Volume Delivered to Zero
1
2
Hold for 3 beeps
3
4
Reset Alarms
1
Correct alarm condition
2
3
Reset Volume Delivered
(current dose) & Dose Limit
to Zero & Infusion Rate to 5
1
2
3
Alarms & Indicators
OCC/EMP
→Pathway occluded
→Container empty
→Drip chamber improperly installed, or
→Pump inactive for over 2½ minutes
DOSE COMP
→Dose limit has been reached.
(4 beeps, flashing indicator,
2 ½ minute repeats)
FLO FAST
→Tubing not on rotor, or
→Drop sensor blocked over 6 seconds.
(Loud 3 tone beep)
LOW BAT
→Low battery. Plug into AC outlet
16
Hold for 7 beeps
EN-US-2017-08-16C