Drawell TD3002 사용자 설명서 - 페이지 5

{카테고리_이름} Drawell TD3002에 대한 사용자 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Drawell TD3002 6 페이지. Precision balance

the weight of the individual sample should not be less than the division of the
balance.
Return to Weighing Mode: Press key PCS to return to Weighing mode.
OVERLOAD
The weight of the objective cannot exceed the rated maximum capacity of the
balance, when exceeding, "-------" will be shown in the display, remove the
objective immediately from the platter so as to prevent damages to the
balance.
TIPS
 Warming up is necessary before any operation to the balance.
 In TARE mode, the value of the tare cannot exceed the maximum capacity
of the balance
 Calibration is necessary to ensure a reliable weighing.
 Switch off the balance when it is not used.
 It is suggested to turn the round platter clockwisely before take it off the balance
Maintenance ----- Trouble shooting and solutions
Phenomenon
Upper Line - - - -
Under Line_ _ _ _
Err-1
Err-2
Possible Reason
Over Load
The object is not properly loaded to
the balance
Too frequently turn on and off the
balance
The balance is not stablised
Solution
Re-calibrate the Balance
Re-calibrate the Balance
Turn off the balance, resume it after 3
seconds
Wait for a few seconds for stabilization