5. Remark:
1)When the box sensor wire break or short circuit, temperature display ER1; Door sensor
wire break or short circuit, temperature display ER2 , Water sensor wire break or short circuit,
temperature display ER3 ,Water level detection error, display ER4,continuous alarm buzzer;
2) Measuring temperature > set temperature + AL, the buzzer alarm intermittently, When
measuring temperature < set temperature + temperature alarm value, the buzzer closed alarm,
press any key cancelable over temperature alarm sound.
Ⅶ. Use
Watering:
1.
Connect the rubber hose with the inlet and outlet which back of the equipment,
the other end connected to the water, open the inlet-outlet valves. Using water pressure to
force water injection water jacket. When the injected water
above the electric heat pipe), can
level lamp (may be water shortage alarm at that time ) until the high water level indicator
light green to stop water, close the inlet-outlet valves.
2.In order to prevent the water jacket in scale, please use pure or distilled water when filled.
3. The configuration of the PH liquid and washing liquid
According to the following mixing PH liquid and washing liquid, and inject in PH and
washing bottles respectively.
A. Washing liquid - copper sulfate solution
1.4g
0.25ml
B. PH liquid - liquid phenol red
changing it about two months
Temperature(℃)
NaHCO3(g)
temperature(℃)
NaHCO3(g)
Ⅷ. Use operations sequence
After incubator installing, follow these steps
1. Open the door, and clean indoor, put shelf.
2. Set to 2/3 water in humidity when need, place it at the bottom of studio and then close the
door.
3. Open the power switch which behind the control box, incubator should be displayed
temperature and WH indicator is bright.
4. Push the "UV" button, sterilization for workshop. (Notice: Please do not click this button
when there are incubation samples inside of the chamber!!!) After the sterilization pressing
the "UV" button to close the germicidal lamp.
5. Set the required temperature first, and then heating incubator, (see the five, the controller
instructions)
Web : www.drawell.com.cn
power-on and continue injected water, viewing water
Redistilled water
0.2% phenol red
NaHCO3 following table
14
16
18
1.87
1.80
1.73
28
30
32
1.43
1.38
1.33
Drawell International Technology Limited
Tel: +86-023-63268643
more than 12L (water level
CuSO2 / 5H2O
Musk toppings
Distilled water
According to the room temperature ,
1000ml
3ml
20
22
1.67
1.61
34
36
1.28
1.23
Email : [email protected]
0.7~
H2SO4
(little)
500ml
24
26
1.54
1.48
37
38
1.21
1.18