Electrical Safety

Lethal electrical hazards can be present in all lasers, particularly in laser
power supplies. Every portion of the electrical system, including the printed
circuit boards, should be considered to be at a dangerous voltage level.
Avoid potential shock by following these guidelines.
Turn off the power switch and unplug the power cord before servicing the
cytometer, unless otherwise noted.
Connect the equipment only to an approved power source. Do not use
extension cords. Have an electrician immediately replace any damaged
cords, plugs, or cables. Refer to the BD LSR II Facilities Requirement
Guide for specific information.
Do not remove the grounding prong from the power plug. Have a qualified
electrician replace any ungrounded receptacles with properly grounded
receptacles in accordance with the local electrical code.
For installation outside the US, use a power transformer or conditioner to
convert the local power source to meet the BD LSR II power requirements
(120 V ±10%, 50/60 Hz). Contact your local BD office for further
information.