BD Sensors DS 6 설치 지침 - 페이지 2

{카테고리_이름} BD Sensors DS 6에 대한 설치 지침을 온라인으로 검색하거나 PDF를 다운로드하세요. BD Sensors DS 6 2 페이지. Pressure switch
BD Sensors DS 6에 대해서도 마찬가지입니다: 운영 매뉴얼 (2 페이지), 운영 매뉴얼 (2 페이지), 운영 매뉴얼 (2 페이지)

CS
EN
Návod k montáži /
Mounting instructions
Tlakový spínač / Electronic Pressure Switch
DS 6
DS 6
www.bdsensors.com
Centrála východní Evropa /
Headquarters Ostern Europe
BD SENSORS s.r.o.
Hradišťská 817
CZ - 687 08 Buchlovice
Česká republika / Czech Republic
Tel.: +420 572 / 411 011
Centrála západní Evropa /
Headquarters Western Europe
BD SENSORS GmbH
BD-Sensors-Str. 1
D - 95199 Thierstein
Německo /
Germany
Tel.: +49 (0) 92 35 / 98 11-0
Fax: +49 (0) 92 35 / 98 11-11
Tento návod k montáži je výtah z podrobného
návodu k obsluze. Ten si prosím stáhněte
z našich internetových stránek, pokud nejste
s výrobkem dobře seznámeni.
These mounting instructions are an excerpt from the complete
operating manual. It may be downloaded from our homepage, if
you are not familiar with the device.
www.bdsensors.cz
www.bdsensors.com
– Technické změny vyhrazeny –
– Technical modifications reserved –
English
WARNING! In order to avoid hazards to operators and
damages to the device, the following instructions have to
be performed by qualified technical personnel.
WARNING! Adhere to the safety and operating instructions
stated in the operation manual. Effective regulations on
occupational safety, accident prevention as well as national
installation
standards
and
approved
engineering
techniques must in addition be complied with.
Limitation of liability
If the instructions in the operating manual are not adhered
to or if the device is inappropriately used, modified or
damaged, liability is not assumed and warranty claims will
be excluded.
Intended use
Ensure that the medium is compatible with the media-
wetted parts and that the device is suitable for the
application without restrictions. The technical data listed in
the current data sheet is binding and must definitely be
observed.
Product identification
type
ordering
serial
input
designation
code
number
supply
set point
pin
with setting
configuration
Fig. 1 manufacturing label DS 6
Pin configuration
M12x1 (4-pin)
M8x1 (4-pin)
Electrical connection
metal
metal
Supply +
1
1
Supply –
3
3
Contact 1
4
4
Contact 2
2
2
Shield
plug housing
Mounting
WARNING! Install the device only in depressurized and currentless state!
WARNING! Use a suitable seal, corresponding to the medium and the pressure input.
WARNING! Before the installation: make sure that the inside of the threaded neck of the closed valve is clean. If need, remove
the residual liquid in a proper way. Do not mount the device against the liquid inside the thread connection.
Connection G1/4" according to DIN 3852
approx. 5 Nm
Wiring diagrams
with 1 contact
supply +
supply -
contact
For the electrical connection a shielded and twisted multicore cable is recommended.
with 2 contacts
supply +
V
S
supply –
R
L
R
L
contact 1
U
contact 2
MA_DS6_C-E_26.10.2022