Dream Catcher ME1310 매뉴얼 - 페이지 16
{카테고리_이름} Dream Catcher ME1310에 대한 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. Dream Catcher ME1310 18 페이지. Antenna and propagation (3d) lab 1 introduction to 3d radiation pattern measurement
S
Calibration
21
1. Set up the connections as shown in Figure 13.
2. On the N9912A front panel, press Mode. Use the softkey to select NA, and then select S
3. Press Cal (button #5) on the N9912A front panel. Use the softkey to select Normalize.
4. Then, select Thru.
5. Select Measure after completing the setup.
6. Upon the completion of Step 5, press Finish to complete the Norminal calibration. The calibration
can be repeated again at any time in the event that the calibration kit was not properly connected
during the calibration process.
ME1310 Antenna and
Propagation (3D)
Figure 13 – Setup for S
http://dreamcatcher.asia/cw
calibration
21
.
21
Lab 1 - 16/18